instrument for lactose test reading
- Light source: DC12V20W halogen lamp
- Optical path system: 8-channel vertical optical path system
- Wavelength range: 340-1000nm; can perform single and dual wavelength measurements
- Standard configuration: 405, 450, 492, 630nm, optional for other wavelengths, up to 15 filters can be loaded
- Filter reading range: 0-4.000Abs
- Linear range: 0-3.000Abs
- Resolution: 0.001Abs
- Indicating value error: ≤±0.02Abs
- Stability: ≤±0.003Abs
- Repeatability: CV≤0.3%
- Channel differences: ≤0.03Abs
- Plate-vibration function: available, 3 vibration speed and time adjustable in 0-255sec
- Display & operation: 7-inch color LCD screen, display of whole plate measurement information, touch screen operation
- Workstation: professional software that can store 500 sets of programs and 100000 sample results. Providing calculation modes such as absorbance, linear equation, logarithmic equation, quadratic equation, cubic equation, absorbance percentage concentration logarithmic equation, spline function, inhibition rate, etc
- Power supply: Power adapter: Input AC100-240V,50/60Hz; Output DC12V5A
- Printer: external thermal printer (optional)
- Net weight: approximately 6Kg
- Dimension: Approximately 400mm (L) * 260mm (W)* 220mm(H)